Main body High-strength engineering plastic rack, light, beautiful and corrosion resistant
-
Scope of application adult, child
-
Gas source O2, N2O: 0.28~0.6MPa
-
Flow meter O2: 0.05~1.0L/min 1.1~10L/min
-
N2O: 0.05~1.0L/min 1.1~10L/min
-
Flow rate of rapid oxygen supply 25~75L/min
-
Low oxygen pressure alarm There will be sound alarm when the oxygen pressure < 0.2MPa
-
Vaporizer it has the function of automatic compensation based on pressure,temperature,and flow rate.
-
The regulation range of evaporator concentration is 0~5 vol%.
-
Among Halothane,Enflurane,Isoflurane and Sevoflurane,
-
one can be chosen for application as required by the customer.
-
Respiratory circuit working mode: all-close, semi-close, semi-open APL:≤12.5 kPa
-
Respiratory bellows bellows for adults, bellows for children tidal volume range:0~1500 ml
-
Part of Ventilator
-
Display mode High-definition 5.7” LCD screen display
-
Ventilation mode IPPV, SIPPV, SIMV, MANUAL
-
Ventilation parameters Tidal volume 50~1500ml
-
Rate 2~99 bpm
-
SIMV 2~20 bpm
-
I:E 2:1~1:8
-
Inspiratory trigger pressure -10~20cmH2O
-
PEEP 2~20cmH2O
-
Pressure limitation 10~60cmH2O
-
SIGH 1.5 times the inspiratory time/100bpm
-
Parameters for ventilation monitoring tidal volume, minute ventilation volume, IPPV rate, SIMV rate,
-
total respiratory rate, I/E, peak pressure of airway, pressure – time waveform,
-
flow – time waveform, PEEP, inspiratory trigger pressure,(average pressure, oxygen concentration, battery capacity)
-
Security Alarm System
-
Oxygen concentration alarm upper limit setting range (21%~99%)
-
low limit setting range (18%~80%)
-
Airway pressure alarm upper limit setting range 5~60 cmH2O
-
low limit setting range 0~50 cmH2O
-
Minute ventilation volume alarm upper limit setting range 1~20 L/min
-
low limit setting range 0~19 L/min
-
Tidal volume upper limit setting range 100~1500ml
-
low limit setting range 0~1400ml
-
Sustained high-pressure alarm it will give alarm when stress have consistently
-
been higher than 2.5 cmH2O
-
(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)
-
Power AC 220V 50Hz (UPS, Storage battery(Mechanical Arms)(End-tidal CO2 monitoring)Specific interface of waste anesthesia gas)